Guidance and manufacturer’s declaration — Electromagnetic
immunity
WARNING
• This model is intended for use by medical personnel in hospitals and for use in the
electromagnetic environment specified below.
The customer or the user of this model should assure that it is used in such an
environment.
• Portable and mobile RF communications equipment should be used no closer than
30 cm (12 inches) to any part of this model, including cables specified by Olympus.
Immunity test
IEC 60601-1-2 (2014)
test level
Compliance level
Electromagnetic environment -
Guidance
Conducted RF
IEC 61000-4-6
3V (150 kHz – 80 MHz)
Same as left
Refer to the guidance in the left column.
6V (ISM band of 150 kHz
– 80 MHz)
Same as left
Refer to the guidance in the left column.
ISM (industry, science, and medical care) band of 6.765 MHz – 6.795 MHz, 13.553 MHz –
13.567 MHz, 26.957 MHz – 27.283 MHz, and 40.66 MHz – 40.70 MHz between 0.15 MHz and
80 MHz
Radiated RF
IEC 61000-4-3
3V/m (80 MHz –
2.7 GHz)
Same as left
Refer to the guidance in the left column.
Proximity magnetic
field from RF
communication
equipment
IEC 61000-4-3
Refer to the table of the
next page.
Same as left
Refer to the guidance in the left column.