2.5 Foot switch cable
15
MAJ-2301 INSTRUCTION MANUAL
Ch.2
2.5
Foot switch cable
No.
Name
Description
1
FOOTSWITCH
connector
Connect this connector to the connector on the foot switch when using it with wired
connection.
2
Index
Connect the cable to the terminal on the foot switch so that this marking is visible on
the top.
3
FOOTSWITCH
marking
Indicates that the FOOTSWITCH connector is to be connected to the foot switch on
this side.
4
MICROSCOPE
marking
Indicates that the MICROSCOPE connector is to be connected to the surgical
microscope on this side.
5
Index
Connect the cable to the footswitch terminal on the surgical microscope so that this
marking is visible on the top.
6
MICROSCOPE
connector
Connect this connector to the foot switch terminal on the surgical microscope.
7
Lock
Used for removing from a foot switch or surgical microscope.
2.6
Pairing stickers
No.
Name
Description
1
Pairing stickers
No.1 – No.10
Attach the sticker matching the pairing indication on the paired foot switch onto the
wireless unit.
Refer to Section 3.6, “Pairing”.
1
2
3
4
6
5
7
1