Maintenance Manual
Test Programs
3-11
Page
Test No./
Test Name
Function Outline
<Mapping parameter check>
This program can be used to adjust the mapping sensor.
Press the [Start] button, and select the item to adjust by using the menu buttons on the liquid crystal
panel.
[M1]: Automatically adjust mapping parameters <Mapping auto adjust>
[M2]: Execute mapping and display results <Mapping result display>
[M3]: Manually input mapping parameters <Input parameter>
[M4]:
Check the mapping sensor <Mapping sensor check>
[M1]: Automatically adjust mapping parameters <Mapping auto adjust>
The mapping parameters can be adjusted automatically.
Using the 4-way button, select the number of the wafer type to adjust or register from the list of wafer
type numbers displayed on the liquid crystal panel, and then press the [Start] button.
Insert the wafers to be set into the lowest and highest slots of a cassette.
Press the [M1] button. The elevator moves down to the lower limit, and then moves up and stops at
the upper limit.
If the loader displays that the mapping data is <Result>: OK, press the [M2] button to register the data.
[M1]: Execute the automatic mapping adjustment <Adjust mapping>
[M2]: Register the results of automatic mapping adjustment <Save>
[M2]: Execute mapping and display results <Mapping result display>
This program can be used to check for normal detection when mapping is executed based on the
registered data.
[M1]: Execute mapping <Mapping>
[M2]: Display results of mapping sensor 1 (center) <Display result sensor1>
[M3]: Display results of mapping sensor 2 (far side) <Display result sensor2> AL120-L/LMB86-180,
LMB8-90
[M4]: Display results of mapping sensor 3 (near side) <Display result sensor3> AL120-L/LMB86-180,
LMB8-90
No. 12
Wafer
mapping
adjustment
[M1]: Execute mapping <Mapping>
Using the 4-way button, select the number of the wafer type to adjust or register from the list of wafer
type numbers displayed on the liquid crystal panel, and then press the [Start] button.
Insert all wafers to be confirmed into the cassette.
Press the [M1] button. The elevator moves down to the lower limit, and then moves up and stops at
the upper limit.
[M2], [M3], [M4]: Display mapping results <Display result>
Press the [M2], M3] and [M4] buttons to display the
results on the liquid crystal panel.
The results displayed include the results of judgments for each slot (Pass/Fail), center position of a
wafer <P>, and wafer thickness <Thick>.
[M1]: Next page <PageUp>
[M2]: Previous page <PageDn>
[M3]: Terminate display <Exit>
Summary of Contents for AL120-6Series
Page 2: ...This page intentionally left blank ...
Page 9: ...Operation Maintenance Manual Introduction Safety Precautions i 7 Page AL120 86 Series Unit mm ...
Page 16: ...Introduction Conformity Standards Operation Maintenance Manual i 14 Page ...
Page 17: ...Operation Maintenance Manual Introduction Conformity Standards i 15 Page ...
Page 20: ...Maintenance Manual Table of Contents C 2 This page intentionally left blank Page ...
Page 56: ...Test Programs Maintenance Manual 3 26 Page This page intentionally left blank ...
Page 64: ...Replacement of Consumables Maintenance Manual 4 8 Page This page intentionally left blank ...