Oki Data CONFIDENTIAL
44871001TH Rev.6
6-5
6. MAINTENANCE MENUS
Item1
Item2
Item3
Item4
Item5
Value
Default
value
ODA
Default
value
OEL
Default
value
JP
Notes
Password
Scanner
Maintenance
Adjust Mech. Adjust ADF
Motor
ADF Keep
Current
10 ~ 1400 [mA]
200 [mA] 200 [mA]
Adjust CIS
Adjust CIS
simple
R continuous
G continuous
B continuous
All continuous
Only for engineering test
Sequentially light the designated RGB colors and check them during calibration
configuration duties.
Move the CIS to the standard position.
In the position moved to, sequentially light R, followed by G and B in the same
manner.
Light each color for approximately 3 seconds.
Display "Testing" during execution?
→
On the panel, display "CIS light testing" and "Cancel."
Check CIS
300dpi
600dpi
---Results Displayed---
CCD_SIG9_WID_H 0x000000
CCD_SIG2_WID_H 0x000000
Only for engineering test
Check CIS maintenance displays the exposure time at each resolution.
When resolution is set, scanner calibration is conducted at that
resolution and the
following results are displayed:
Red-1
Red-2 (No indication)
Green-1
Green-2 (No indication)
Blue-1
Blue-2 (No indication)
Lsync
Note : CIS exposure time varies with resolution.
Set CIS
Exposure
time
0 ~ 4294967295
Only for engineering test
Change LED exposure time settings, and then read the document using
PC Scan.
Reading implemented for the LED exposure time in the settings, without
conducting
calibration.
A warning is displayed when a value greater than the Lsync cycle has
been set.
Settings are shown below.
Red-1
Red-2 (No indication)
Green-1
Green-2 (No indication)
Blue-1
Blue-2 (No indication)
Lsync
Summary of Contents for MB461MFP
Page 1: ...MB461 MB471 MB491 MB491 MB491 LP MPS4200mb MPS4700mb Maintenance Manual 031314C ...
Page 72: ...Oki Data CONFIDENTIAL 44871001TH Rev 6 3 21 3 MFP INSTALLATION Sample MB471 Sample MB491 ...
Page 114: ...44871001TH Rev 6 4 36 Oki Data CONFIDENTIAL 4 REPLACEMENT OF PARTS ...
Page 138: ...Oki Data CONFIDENTIAL 5 LUBRICATING POINT 5 1 Lubrication point 5 2 ...