CHAPTER 1
GETTING STARTED
7000 Series T71 Indicator Service Manual
1-19
Ohaus Corporation www.ohaus.com
1.9.2 Interface
Commands
The RS232 Interface enables a computer to be used to control some functions of the Indicator
using the commands listed in Table 1-4.
TABLE 1-5. SERIAL INTERFACE COMMAND TABLE.
Command
1
Function
ON
Turns the indicator on.
OFF
Turns the indicator off.
IP
Immediate Print of displayed weight (stable or unstable).
P
2
Print stable displayed weight (according to stability setting).
CP
Continuous Print
SP
Print when stable.
xP
Interval Print x = Print Interval (1-3600 sec)
Z
2
Same as pressing Zero button
T
2
Same as pressing Tare button
xT
Download Tare value in grams (positive values only). Sending 0T clears tare
(if allowed)
PU
Print current unit: g, kg, lb, oz, lb:oz
xU
Set scale to unit x: 1=g, 2=kg, 3=lb, 4=oz, 5=lb:oz
PV
Version: print name, software revision and LFT ON (if LFT is set ON).
H x “text”
Enter Header line , where x = line number 1 to 5, “text” = header text up to 24
alphanumeric characters
Esc R
Global reset (All menu settings are reset to the original factory defaults.)
Notes
:
1. Commands sent to the Indicator must be terminated with a carriage return (CR) or
carriage return-line feed (CRLF).
2. Alternate command characters may be defined by the user.
3. Data output by the Indicator is always terminated with a carriage return-line feed (CRLF).