CHAPTER 1
GETTING STARTED
7000 Series T71 Indicator Service Manual
1-17
Ohaus Corporation www.ohaus.com
CAL
CAL TEST
tESt
kg
0
tESt
kg
10
tESt
kg
--T--
10.000
kg
0.001
kg
DIFFERENCE
tESt
kg
--T--
1.8.5 Calibration
Test
NOTE
: Calibration Test is always available (even when LFT is set to ON).
Calibration test is used to compare a known calibration weight against the stored span
calibration data. Go to the Calibration menu, and Press
No
until CAL
CAL TEST
appears.
With no load on the scale, press
Yes
. The display shows the zero load
calibration point.
With no load on the scale, press
Yes
. The display shows
--T--
, followed
by the full load calibration point.
Place the specified full load calibration weight on the scale and
press
Yes
. The display shows
--T--
, followed by difference between
the calibration weight and the stored calibration data.
After five seconds, the display returns to the current application mode.
1.8.5 Geographical Adjustment Factor
The Geographical Adjustment Factor (GEO) is used to compensate for variations in gravity.
To determine the GEO factor that corresponds to your location, see Appendix A.
Note:
Changing the GEO Factor alters the calibration. The GEO value was set at the factory
and should only be changed if the location of use varies from the default setting.
To adjust the GEO Factor, go to the Calibration menu, and Press
No
until CAL
GEO
appears.
Press
Yes
. The default GEO setting appears. Press
No
to increment the number, or
Back
to
decrement, until the desired number appears. Then press
Yes
.
Settings from 1 to 31 are available.
CAL
GEO
appears again. Press
No
to exit the
GEO
menu. Then press
Exit
to return to the
current application mode.