CHAPTER 1
GETTING STARTED
Ohaus Corporation www.ohaus.com
1-6
7000 Series T71 Indicator Service Manual
1.6.2 Jumper
Connections
When a 4-wire load cell is attached to terminal block J4, jumpers W1 and W2 must be shorted.
When a 6-wire load cell is attached to terminal block J4, jumpers W1 and W2 must be open.
When a 4-wire load cell is attached to terminal block J7, jumpers W3 and W4 must be shorted.
When a 6-wire load cell is attached to terminal block J7, jumpers W3 and W4 must be removed.
+EXC +SIG -SIG -EXC
+SENS GND -SENS
Figure 1-5. Load Cell wiring.
Pin Connection
1 +Excitation
2 +Sense
3 +Signal
4 GND
5 –Signal
6 –Sense
7 –Excitation
Figure 1-6. Open/shorted jumpers.
1.6.3 RS232 Interface
The T71P has a built-in RS232 Interface: a standard 9-pin D-style female connector on the back
of the T71P Housing. In addition, each model (T71P and T71XW), has an optional COM2
connector, which can be either RS232 or RS422/485. The optional interface is available
separately, with its own instructional manual for installation and configuration.
The RS232 pin configuration for the T71P is illustrated in Figure 1-7, below.
1
2
3
4
5
6
7
8
9
Figure 1-7. RS232 Pins.
Pin Connection
1 N/C
2 TXD
3 RXD
4 N/C
5 GND
6 N/C
7 CTS
8 RTS
9 N/C
For software setup of the RS232 Interface, through the Indicator’s menu, see Section 1.9.
All other aspects of hardware configuration and setup are described in the Instruction Manual.