OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com
20
• Step – This will increase the temperature or pressure up to the
target as fast as possible. Enter the target in the "Target" box.
• Dwell – This will hold the current temperature or pressure for a set
number of minutes. Enter the time (hh:mm:ss) into the “Time” Box.
The maximum temperature setpoint allowed is 600°F (315.6°C). The
unit allows pressure setpoints up to 40,000 psi (275.8 MPa). The units of
measure will be the same as entered in the Setup menu (see page 17).
5. To edit an existing step, double click the step in the step list or highlight
the existing step and select the “Edit” button. Click OK when done.
6. Click the “OK” button to exit the “Setup” screen. The new Temperature
Profile will now appear in the “Available Profiles” list on the “Profile” tab.