Calibration
Temperature
The thermocouple will require calibration quarterly or if it is replaced.
1. Place the thermocouple in a dry block calibrator.
2. Start the calibration procedure:
a. For onboard control, touch Calibrate and then Temperature.
b. For software control, select Temperature from the Calibrate menu
3. Follow the instructions on the screen.