OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com
12
6.
To fill the oil reservoir, remove the right-hand side panel and remove the
oil reservoir cap. Pour approximately four liters (or until full) of oil into the
reservoir. Replace the cap. Make sure the seal is air tight. Use the sight
glass on the front to check the oil level.
Mineral Oil (#120-001) is recommended for tests up to 400°F. For
tests above 400°F, use only Phenylmethlsiloxane Silicone Fluid
(#120-004).
7. Turn the “Main Power” switch On.
8. Calibrate the potentiometer before connecting the Consistometer to a
computer. See page 51.
Reservoir
Cap