EX Series Operating Manual
6-9. ADC SAMPLING
▌Be able to set the output voltage/current measuring speed and displaying speed.
▌If the frequency is lower, data acquisition is reduced. But you can get the more accurate value.
▌If the frequency is higher, data acquisition is increased. But voltage/current measuring value
can be changeable frequently in the display.
▌In case of PC interface control, recommends ADC sampling speed is 20~50Hz.
▌I
f F
t
l
t l
d ADC
li
d i 5 20H
▌In case of Front-panel control, recommends ADC sampling speed is 5~20Hz.
▌CAUTION : If you have to acquire the voltage/current in high speed100Hz, 300Hz, use 1.3KHz
speed!!
▌Function.
> 5Hz
Measures voltage/current at every 200ms.
> 20Hz
Measures voltage/current at every 50ms.(Instrument default value)
> 50Hz
Measures voltage/current at every 20ms.
> 100H
Measures voltage/current at every 10ms
> 100Hz
Measures voltage/current at every 10ms.
> 300Hz
Measures voltage/current at every 3.3ms.
> 1.3KHz
Measures voltage/current at every 770us.
ADC SAMPLING
▌Power switch on in a pressing "FACTORY" key.
▌
p
g
y
▌Power switch ON
▌When appearing "1. user-mem clear" message, release "FACTORY" key.
In a pressing
Release
▌Enter to "7. adc sampling" mode by using cursor key.
▌In other to set, press the "FACTORY" key one more time.
LCD Display
▌ Set the "50Hz" mode by using cursor key
20hz
▌ Set the 50Hz mode by using cursor key.
LCD Display
▌To finish the "ADC SAMPLING" mode, press the "FACTORY" key
one more time.
LCD Display
done
50hz
≫ Related remote interface command
FACT:ADC {5|20|50|100|300|1300}
FACT:ADC?
Return Value(5Hz, 20Hz, 50Hz, 100Hz, 300Hz, 1.3KHz)
Apply:
FACT:ADC 100
Set 100Hz speed in other to acquire the data.
FACT:ADC?
Return value : 100Hz
Confirm the ADC sampling speed : 100Hz.
- 81 -