EX Series Operating Manual
3-13. Recall
You can recall up to ten different operating states(Output ON/OFF state, Voltage, Current, OVP,
OCP, UVL, OVL, UCL, OCL) in non-volatile memory.
The following steps show you howto recall an operating state.
Recall Key can be used to FACTORY function.
Recall Key can be used to FACTORY function.Refer to "5.FACTORY ".
▌Turn on the power supply
After turned on, check the power supply displays"
**OUTPUT OFF**
" Mode.
▌Press the "RECALL" key in other to recall the stored state.
LCD Display
RECALL NO, 01
▌Choose the storage location number that you want to apply by using
encoder knob. For examples, if you want to apply No.02 state,
turn to right one click.
LCD Display
RECALL
NO, 02
LCD Display
RECALL NO, 02
▌In other to recall in your selected memory location 2,
press the "RECALL" key.
LCD Display
saved
▌Appear "Saved" message, and return to previous state.
≫ Related Remote Interface Command
*RCL {1|2|3|4|5|6|7|8|10}
Apply :
*RCL 4
recall operating states at location 4 in non-volatile memory.
If you store operating state when the power supply output is on, it may give rise to serious trouble
when you recall. So. It is safe to store under the output off mode.
Note
- 49 -