13 Cleaning, di
s
infection and maintenance
40 / 60
Instruction Manual Easyfield® (GA/15000/XXXX/EN 1219 Rev01)
13 Cleaning, di
s
infection and maintenance
This section describes how to clean and disinfect the Easyfield®.
Sterilization is not required.
Heed the product descriptions and instruction manuals of products and
equipment you use to care for, clean, and disinfect the unit and/or its
accessories.
Warning
Risk or personal injury during service or maintenance works
If you use the device during service or maintenance works (cleaning, dis-
infection and maintenance) personal injury may occur.
Do not use the device during service or maintenance works.
13.1 Cleaning
Caution
Risk of electric shock if the Easyfield® is not completely disconnected
from the mains for these jobs.
Turn the Easyfield® off,
Pull the power plug before cleaning. When disconnecting electrical
connections, pull on the respective plug and not on the cable itself.
Do not clean the Easyfield® with aggressive, chlorine-containing,
abrasive or sharp cleaning agents.
Required materials:
Anti-static cleaner for plastic surfaces
Cleaner for painted surfaces: mixture of equal parts mineral spirits
and distilled water, with a few drops of household detergent if
needed
Soft cloth or lens brush
Alcohol or lens cleaner
Cleaning inervalls
Clean the chin support and forehead rest after each examination; clean
the housing as needed.
Clean the hand held button after each examination.