4 Safety in
s
truction
s
10 / 60
Instruction Manual Easyfield® (GA/15000/XXXX/EN 1219 Rev01)
Caution
Use of a multiple socket extension cord
Risk of personal injury or material damage caused by unsafe multiple
socket extension cord
If you use a multiple socket extension cord to connect the
Easyfield® to the power supply, you must heed the following information:
Use an extension cord that complies with the requirements of DIN EN
ISO 60601-1: 20005, section 16.
Do not place the multiple socket extension cord on the floor.
Do not use more than one multiple socket extension cord.
Plug only the Easyfield® and the computer that is being used with the
unit (if applicable) into the multiple socket extension cord.
If you are using a multiple socket extension chord it has to be supplied
with a isolation transformer.
If you are using a new computer for the Easyfield®, you must have the
electrical safety checked. Call OCULUS Service for this purpose.
Electromagnetic Compatibility (EMC/Cables)
Risk of personal injury or damage to property due to electromagnetic in-
terference
Portable and mobile RF communications equipment can affect medical
electrical equipment
.
Make sure that portable and mobile RF communications equipment do
not cause interference.
Recommendation: Maintain a minimum distance of 4 m. If the
distance is shorter, you must ensure that the Easyfield® functions
correctly.
Cybersecurity
☞
Do not use the Easyfield® with wireless technology, for example with
wireless USB
To ensure cyber security in order to the usage of the device, the following
security measures should be considered. Contact your computer
administrator: