19 Annex
52 / 60
Instruction Manual Easyfield® (GA/15000/XXXX/EN 1219 Rev01)
To be in compliance with the requirements of the IEC 60601-1-2. 6.1 and
6.2 the following types of equipment, accessories, power adapters and
cables must be used.
19.2 Guidance and Manufacturer’
s
Declaration -
Electromagnetic Emmi
sss
ion
s
and Immunity for the Ea
s
yfield®
Order number
De
s
cription
15000
OCULUS-Easyfield® C
15005
OCULUS-Easyfield® S
05200320
Cable with connector plug,
EU standard
2.5m
05200210 ( 110 Volt)
Cable with connector plug,
US standard
2.5m
015692000010
USB FS Med isolator
05150805
Power supply
GSM40B12-P1J
Guidance and manufacturer´
s
declaration electromagentic emmi
ss
ion
s
IEC 60601-1-2, 5.2.2.1, table 1
The OCULUS Easyfield® is intended for operation in the electromagnetic environment specified below. The user
of the Easyfield® should ensure that it is being used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The Easyfield® uses RF energy only for its internal func-
tion. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic
equipment.
HF-emissions
CISPR 11
Class B
Harmonics emissions
IEC 61000-3-2
Class A
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
complies