s
Instruction Manual Easyfield® (GA/15000/XXXX/EN 1219 Rev01)
9 / 60
Instructions for operation
Before first use: Let OCULUS or an authorized dealer train you in the
operation of the Easyfield®.
Never operate a damaged Easyfield®.
Only operate the Easyfield® using original accessory parts supplied by us,
and when the device is in technically correct working order.
Do not touch the patient and the Easyfield® simultaneously.
Make sure that the device cannot tip over by leaning against it or sitting
on it.
Do not use the device if you have not understood the Instruction
Manual.
Notes on maintenance
When cleaning, use a damp cloth and make sure that no liquid enters
the Easyfield®.
To ensure that it functions correctly and safely we recommend the
following: Have the Easyfield® checked every two years by our service
department or an authorized dealer. If an error occurs which you are
unable to correct, label the Easyfield® as "out of order" and contact our
service department or an authorized dealer.
Instructions on disassembly and disposal
When disconnecting electrical connections, pull on the respective plug
instead of the cable itself.
Dispose of the unit in conformance with legal requirements.
Instructions on Electrical Safety
Caution
Risk of personal injury or damage to property due to an incorrect level of
safety
Connecting the Easyfield® with its non-medical electrical equipment (e.g.
data processing equipment) to a medical electrical system must not result in
a patient safety level below that prescribed by DIN EN ISO 60601-1. If making
this connection leads to the leakage current threshold being exceeded,
protective measures including a circuit breaker must be in place.
Ensure that connections with non-medical devices are made correctly.
Only use the power adapter listed in the packing list.
Use only a computer that meets the specifications given in this
instruction manual,
.