OM0283
Page 47 of 70
Rev 2 July/2021
Table 7.0
Recommended Measurement Methods for Cabinet Downflow and Inflow
A. Downflow Measurement
a. Recommended Instruments: TSI 8355 Thermo anemometer
b. Procedure: Supply filter efflux is measured on a grid, in a horizontal plane 4 inches (102mm) above the bottom edge
of the window. No reading should be taken closer than 6 inches (152mm) from the inside perimeter.
c. Test Data - Inches (mm):
Model Size
Window Access Height
300
10/8/12
6.25
(159)
13.542
(344)
20.834
(529)
28.125
(714)
400
10/8/12
7.0
(178)
12.396
(315)
17.792
(452)
23.188
(589)
28.584
(726)
33.980
(863)
39.375
(1000)
500
10/8/12
7.25
(184)
13.518
(343)
19.786
(503)
26.054
(662)
32.322
(821)
38.590
(980)
44.858
(1139)
51.125
(1229)
600
10/8/12
7.5
(191)
15.410
(391)
23.320
(592)
31.230
(793)
39.140
(994)
47.050
(1195)
54.960
(1396)
62.870
(1597)
6.25
(159)
7.0
(178)
7.250
(184)
7.5
(191)
12.5
(318)
12.5
(318)
12.5
(318)
12.5
(318)
18.750
(476)
18.0
(457)
17.750
(451)
17.5
(445)
Number of Readings:
Average Velocity
ft./min.(m/s)
d. Acceptance Criteria:
1. Average downflow velocity = 55 to 65 fpm (.28 to .33 m/s)
2. Individual readings must be 20% or + 16fpm (+ 0.08m/s) whichever is greater (factory test) or
+ 25% or + 16fpm (+ 0.08m/s) whichever is greater (field test) from the average downflow velocity.
B. Inflow Measurement
a. Recommended Instrument: Shortridge Flowhood ADM-870 or TSI 8355 Thermo anemometer.
b. Primary Procedure:
The primary procedure to determine inflow velocity uses a Direct Inflow Measurement (DIM) Instrument (i.e.
Shortridge flowhood). The DIM Instrument can be used directly on the cabinet with NO CORRECTION FACTORS
REQUIRED if operated in the local density default mode. NSF has tested the cabinet and established listed air
velocities. Use duct tape to secure the DIM Instrument to the cabinet preventing any sneak air paths from occurring.
The DIM Instrument will read inflow volume (i.e. CFM). Use the window access opening area to calculate inflow
velocity.
Summary of Contents for LabGard D NU-543-400
Page 7: ...OM0283 Page 7 of 70 Rev 2 July 2021 ...
Page 9: ...OM0283 Page 9 of 70 Rev 2 July 2021 ...
Page 10: ...OM0283 Page 10 of 70 Rev 2 July 2021 ...
Page 11: ...OM0283 Page 11 of 70 Rev 2 July 2021 ...
Page 12: ...OM0283 Page 12 of 70 Rev 2 July 2021 ...
Page 16: ...OM0283 Page 16 of 70 Rev 2 July 2021 ...
Page 21: ...OM0283 Page 21 of 70 Rev 2 July 2021 ...
Page 34: ...OM0283 Page 34 of 70 Rev 2 July 2021 ...
Page 65: ...OM0283 Page 65 of 70 Rev 2 July 2021 ...
Page 66: ...OM0283 Page 66 of 70 Rev 2 July 2021 ...
Page 67: ...OM0283 Page 67 of 70 Rev 2 July 2021 ...
Page 68: ...OM0283 Page 68 of 70 Rev 2 July 2021 ...