6.1.2
Front Panel
The control system front panel contains the following functions described in detail (see Drawing BCD-16377).
6.1.2.1 Blower Key
The blower key controls the ON/OFF power to the blower.
LED above key indicates: Full green for blower on
Blinking green for blower pending
Blinking full red for blower alarm
6.1.2.2 Hidden Key
The hidden key is located just above the blower LED indicator centered in the airflow symbol. The hidden key
is used for various functions including the blower password 3 key sequence if the option is activated.
6.1.2.3 Fluorescent Light Key
The fluorescent light key controls the on/off power to the fluorescent light.
LED above the key indicates full blue for fluorescent light on, blinking blue during auto timer duration.
6.1.2.4 Ultraviolet (UV) Light Key
The UV light key controls the on/off power to the UV light (if optionally installed).
LED above indicates full yellow for UV light on, blinking yellow during auto timer duration.
6.1.2.5 Outlet Keys
The outlet key controls the ON/OFF power to the outlets.
LED above indicates full blue for outlets on, blinking blue during auto timer duration.
6.1.2.6 Red Alarm LED
The red alarm LED will indicate any alarm condition and remain indicating until the alarm condition is cleared.
A blinking red alarm LED indicates a power interruption has occurred.
6.1.2.7 Audible Alarm Silence Key
The audible alarm silence key allows user interaction to silence an audible alarm for a period of 15 minutes.
After 15 minutes if the alarm condition still exists, the audible alarm will again sound.
The audible alarm silence key also is used to exit all
FlowGard™
user interaction menus.
6.1.2.8 Arrow Adjustment Keys
The arrow adjustment keys allow user interaction for various functions.
6.1.2.9 ENT Key
The ENT key is used to select item on display or accept and enter current item, function or value.
6.1.3
Control System Power
After the LabGard® ES NU-543 is plugged into the appropriate facility line power the control system will power up
and display a power on reset message, the last revision of software (for several seconds) and time. The control
panel will also indicate the power up status by blinking the red alarm LED. Pressing any key will acknowledge the
power up status and turn off the blinking red alarm LED.
If a power interruption occurs, all control system functions, calibrations and parameters will be maintained and
continue upon restoration of power. Just as the initial power up, the red alarm LED will blink to indicate power up
status along with the display message of power on reset.
Airflow idle XX:XX
GAF3 Version XX:XX
Power on Reset
Summary of Contents for LabGard D NU-543-400
Page 7: ...OM0283 Page 7 of 70 Rev 2 July 2021 ...
Page 9: ...OM0283 Page 9 of 70 Rev 2 July 2021 ...
Page 10: ...OM0283 Page 10 of 70 Rev 2 July 2021 ...
Page 11: ...OM0283 Page 11 of 70 Rev 2 July 2021 ...
Page 12: ...OM0283 Page 12 of 70 Rev 2 July 2021 ...
Page 16: ...OM0283 Page 16 of 70 Rev 2 July 2021 ...
Page 21: ...OM0283 Page 21 of 70 Rev 2 July 2021 ...
Page 34: ...OM0283 Page 34 of 70 Rev 2 July 2021 ...
Page 65: ...OM0283 Page 65 of 70 Rev 2 July 2021 ...
Page 66: ...OM0283 Page 66 of 70 Rev 2 July 2021 ...
Page 67: ...OM0283 Page 67 of 70 Rev 2 July 2021 ...
Page 68: ...OM0283 Page 68 of 70 Rev 2 July 2021 ...