OM0283
Page 32 of 70
Rev 2 July/2021
6.6.3
Assembly
a. Replace front grill (if removed).
b. Replace the work tray and carefully tighten the thumbscrews.
c. Replace perforated metal diffuser screen over the underside of the supply HEPA filter.
d. Wipe down all exposed surfaces of the work area with 70% isopropyl alcohol.
e. Prepare for aseptic operation.
¹ Available from Lab Safety Supply, Janesville, WI 53547-1368, or other laboratory, industrial, or hospital supply distributors.
²American Society of Hospital Pharmacists. 2006. ASHP Guidelines on Handling Hazardous Drugs
Am. J. Hosp. Pharm.
63:1172-1193.
7.0
General Maintenance
All maintenance actions on this equipment must be performed by a qualified
technician who is familiar with the proper maintenance procedures required
including both certification and repair.
7.1 Decontamination
No maintenance should be performed on the interior of the LabGard® ES cabinet (area behind access panels) unless the
cabinet has been microbiologically decontaminated, is known to be biologically clean, or known to be chemically inert.
Surface disinfection is performed as specified in the Cleaning Procedures section.
Hazardous Gases! Personal Protection Equipment Required.
Decontamination using hazardous gas, vapor or mist
must be performed in accordance with
the specifications of NSF/ANSI 49, Annex G, EN12469, Annex J or applicable national, state, province
or local regulations
This procedure presents considerable risks and must be performed only by specially trained and authorized service
personnel in accordance with applicable national safety regulations.
The decontaminate is generated either external or internal of the sealed cabinet. The quantity of decontaminate should
follow standard or manufacturer’s recommendations based on cabin
et volume. The decontaminate process should follow
standards or manufacturer’s recommendations based on the decontaminate used.
All decontaminate materials are hazardous (chemical-liquid, gas and vapor)
(Flammable
–
process) and are required to be handled properly. Follow all
product and process documentation and labelling.
CAUTION
!
CAUTION
!
CAUTION
!
Summary of Contents for LabGard D NU-543-400
Page 7: ...OM0283 Page 7 of 70 Rev 2 July 2021 ...
Page 9: ...OM0283 Page 9 of 70 Rev 2 July 2021 ...
Page 10: ...OM0283 Page 10 of 70 Rev 2 July 2021 ...
Page 11: ...OM0283 Page 11 of 70 Rev 2 July 2021 ...
Page 12: ...OM0283 Page 12 of 70 Rev 2 July 2021 ...
Page 16: ...OM0283 Page 16 of 70 Rev 2 July 2021 ...
Page 21: ...OM0283 Page 21 of 70 Rev 2 July 2021 ...
Page 34: ...OM0283 Page 34 of 70 Rev 2 July 2021 ...
Page 65: ...OM0283 Page 65 of 70 Rev 2 July 2021 ...
Page 66: ...OM0283 Page 66 of 70 Rev 2 July 2021 ...
Page 67: ...OM0283 Page 67 of 70 Rev 2 July 2021 ...
Page 68: ...OM0283 Page 68 of 70 Rev 2 July 2021 ...