OM0283
Page 40 of 70
Rev 2 July/2021
a. To install the supply filter, simply reverse the procedure outlines in Step 3a, above.
Note:
Be sure to open the choke plate fully before inserting the filter into the tray.
This will assist in adjusting the airflow.
b. To install the exhaust filter, apply a thin layer of silicone grease to the top and bottom
gaskets of the filter and carefully insert into the exhaust choke tray.
Position the filter frame within the outside walls of the exhaust opening on the top of the hood.
Tighten the spring loaded bolts, 4 places, depressing the gasket material by 1/8 inch (3mm)
Step 5: Motor/Blower Assembly Removal
a. It is recommended that the motor/blower to be removed as a single unit. To remove,
disconnect electrical connections to the motor, remove the HEPEX pressure plenum and
unbolt the motor/blower assembly from the roof of the cabinet (4 places). Always inspect
the rubber isolation motor mounts and replace those that are cracked or visibly show stress.
b. Replace the motor exactly as originally installed in the blower housing, paying particular
attention to the correct electrical connections (see Electrical Schematic).
c. Re-install the new motor/blower assembly.
Summary of Contents for LabGard D NU-543-400
Page 7: ...OM0283 Page 7 of 70 Rev 2 July 2021 ...
Page 9: ...OM0283 Page 9 of 70 Rev 2 July 2021 ...
Page 10: ...OM0283 Page 10 of 70 Rev 2 July 2021 ...
Page 11: ...OM0283 Page 11 of 70 Rev 2 July 2021 ...
Page 12: ...OM0283 Page 12 of 70 Rev 2 July 2021 ...
Page 16: ...OM0283 Page 16 of 70 Rev 2 July 2021 ...
Page 21: ...OM0283 Page 21 of 70 Rev 2 July 2021 ...
Page 34: ...OM0283 Page 34 of 70 Rev 2 July 2021 ...
Page 65: ...OM0283 Page 65 of 70 Rev 2 July 2021 ...
Page 66: ...OM0283 Page 66 of 70 Rev 2 July 2021 ...
Page 67: ...OM0283 Page 67 of 70 Rev 2 July 2021 ...
Page 68: ...OM0283 Page 68 of 70 Rev 2 July 2021 ...