OM0283
Page 33 of 70
Rev 2 July/2021
7.1.1
Manual Decon
1. Disconnect power to the cabinet. Remove screws at each upper side of the control center and allow the
control center to rotate down, resting on the safety straps. Disconnect electrical connectors on left side.
Disconnect electrical from right side. (Be sure to note the location of the supply and exhaust sensor wires
before disconnecting them from the main board). Remove control center by disconnecting safety straps
lifting the control cable up and away from cabinet. Fold and pinch tubing to seal.
2. Remove the front decorative panel via top/front fasteners.
3. Remove window assembly (both frame and window, either being manual, sliding, or automatic) via the
upper and lower fasteners (Refer to BCD-16376). At this point, the sliding window assembly can be
removed.
4. Remove exhaust sensor shrouds.
5. Remove the diffuser screen and gasket around perimeter of workzone (if present).
6. Prepare decontamination equipment. Reference decontamination procedure and use the following chart
to calculate chemical requirements.
Note: The outlets in the work area are energized as long as the cabinet is plugged in and
switched on the front panel. Unplug the cabinet before decontamination equipment is
plugged into these outlets. The control centers electrical connectors may be re-attached
to utilize the cabinet's interior outlets and fan during the decontamination.
8. Set control center on its side resting on the left side of cabinet. Then plug in the 16-pin connector cable
from the cabinet to the control center. Reconnect power to the cabinet. The outlets and blower can then
be activated.
9. Seal front and top openings using plastic and tape.
BE SURE CABINET IS TOTALLY SEALED TO PREVENT ANY
LABORATORY EXPOSURE TO DECONTAMINATION GAS,
VAPOR OR MIST.
10. Perform decontamination procedure.
7.1.2
Auto Decon
The Auto Decon process is intended to be used with any automated cabinet decontamination process (i.e
hydrogen peroxide, chlorine dioxide, etc.) The automated decon process parameters are typically set up
independently of the cabinet. Inlet and outlet ports from the automated decon machine are connected to the
cabinet. The cabinet control system may be set up to run simultaneously aiding the decontaminate circulation
throughout the cabinet during the entire decon cycle.
To validate two decon processes
and associated cabinet set up procedure; NuAire utilized a Bioquell Clarus II H₂O₂
and a Steris X10 the following process was performed by the Bioquell Clarus II.
1.
Remove armrest via fasteners.
2.
Remove exhaust shroud and sensor.
3.
Seal top and front openings using either heavy plastic and duct tape or seal plates.
Seal all joints with tape as indicated on drawing BCD-16509.
Note:
Top and front seal plates may be obtained by contacting NuAire representative or distributor.
4.
Attach inlet and outlet connections to decontamination system along with pressure and/or concentration
sensor if provided.
Cabinet Size
300
400
500
600
Cabinet
Dimensions
58-7/8 x 28 x 34-3/8 in.
(1.50 x .711 x .873 m)
58-7/8 x 28 x 46-3/8
(1.50 x .711 x 1.18 m)
58-7/8 x 28 x 58-3/8
(1.50 x .711 x 1.48m)
58-7/8 x 28 x 70-3/8
(1.50 x .739 x 1.88 m)
Cabinet
Volume
32.79 cu. ft.
(.929 cu. m)
44.24 cu. ft.
(1.25 cu. m)
55.69 cu. ft.
(1.58 cu. m)
67.14 cu. ft.
(1.90 cu. m)
CAUTION
!
Summary of Contents for LabGard D NU-543-400
Page 7: ...OM0283 Page 7 of 70 Rev 2 July 2021 ...
Page 9: ...OM0283 Page 9 of 70 Rev 2 July 2021 ...
Page 10: ...OM0283 Page 10 of 70 Rev 2 July 2021 ...
Page 11: ...OM0283 Page 11 of 70 Rev 2 July 2021 ...
Page 12: ...OM0283 Page 12 of 70 Rev 2 July 2021 ...
Page 16: ...OM0283 Page 16 of 70 Rev 2 July 2021 ...
Page 21: ...OM0283 Page 21 of 70 Rev 2 July 2021 ...
Page 34: ...OM0283 Page 34 of 70 Rev 2 July 2021 ...
Page 65: ...OM0283 Page 65 of 70 Rev 2 July 2021 ...
Page 66: ...OM0283 Page 66 of 70 Rev 2 July 2021 ...
Page 67: ...OM0283 Page 67 of 70 Rev 2 July 2021 ...
Page 68: ...OM0283 Page 68 of 70 Rev 2 July 2021 ...