OM0283
Page 43 of 70
Rev 2 July/2021
7.5.4
Airflow Calibrations
Failure to calibrate airflow to the specified requirements may
result in unsafe conditions of performance (i.e. product and/or
personnel protection, noise and vibration)
The NU-543 airflow calibration consists of adjustments to balance the airflow within the cabinet.
THIS WORK
SHOULD BE DONE ONLY BY A QUALIFIED TECHNICIAN WHO CAN MEASURE THE AIRFLOW FROM THE FILTERS
WITH A SUITABLE VELOMETER.
NuAire provides two adjustments to balance the airflow within the cabinet.
These are:
a. PWM signal adjust via calibration parameter menu
b. exhaust filter choke
The PWM signal or blower speed adjustment establishes the motor speed controls curve starting point of the
programmed internal reference performance curve while the choke adjusts or balances the exhaust airflow as well
as makes up for filter resistance tolerances. Since it has been NuAire's experience that the filters may not "load"
evenly, choke adjustments may be necessary for proper cabinet airflow balance and performance.
DC ECM motor PWM signal DC voltage should also be monitored and recorded upon final calibration. The DC
voltage may be measured using a digital voltmeter. The two test points to measure DC ECM motor voltage are
located on the DC motor connector on the main control board.
The cabinet is considered to be certifiable if the following airflow measurements are present:
a. Downflow average: 60 LFPM
5 LFPM (.30 m/s
.025 m/s).
b. Inflow average:
105 LFPM
5 LFPM (.53 m/s
.025 m/s) using the direct inflow measurement
method or constricted 3 inch (76mm) high access opening measurement method.
Note:
It is recommended to perform the airflow calibration process at least annually to optimize both the airflow
control and IntelFlow™ monitor systems.
CAUTION
!
Summary of Contents for LabGard D NU-543-400
Page 7: ...OM0283 Page 7 of 70 Rev 2 July 2021 ...
Page 9: ...OM0283 Page 9 of 70 Rev 2 July 2021 ...
Page 10: ...OM0283 Page 10 of 70 Rev 2 July 2021 ...
Page 11: ...OM0283 Page 11 of 70 Rev 2 July 2021 ...
Page 12: ...OM0283 Page 12 of 70 Rev 2 July 2021 ...
Page 16: ...OM0283 Page 16 of 70 Rev 2 July 2021 ...
Page 21: ...OM0283 Page 21 of 70 Rev 2 July 2021 ...
Page 34: ...OM0283 Page 34 of 70 Rev 2 July 2021 ...
Page 65: ...OM0283 Page 65 of 70 Rev 2 July 2021 ...
Page 66: ...OM0283 Page 66 of 70 Rev 2 July 2021 ...
Page 67: ...OM0283 Page 67 of 70 Rev 2 July 2021 ...
Page 68: ...OM0283 Page 68 of 70 Rev 2 July 2021 ...