CALIBRATION of the x0.1 RANGE
LOW VOLTAGE / LOW NOISE HEADSTAGE
(SEC SYSTEMS with Adapted Amplifier)
This headstage has an output current range of
15nA into maximum 100 M
, and the noise and bias current
are reduced by a factor of 10 compared to the standard headstage. Therefore, it is recommended for whole
cell patch clamp recordings.
Current related signals:
CURRENT DISPLAY: X.XXX pA (100 = 100 pA)
HOLDING CURRENT: X.XX nA (100 = 1 nA)
Input sensitivity: BNC labeled
1 nA/V
i.e.
1V = 1nA
BNC labeled
0.1 nA/V
.
1V = 100 pA
)
Output sensitivity: 1V/nA-100V/nA, with 1 V/nA (= 1 mV/pA) 100 pA is 100 mV.
BRIDGE BALANCE: standard as indicated at the front panel (100 = 100 M
)
R
EL
mode display: standard as indicated at the front panel (100 = 100 M
)
The
potential
input and output signals are
not
affected.