CALIBRATION of the x0.1 RANGE
LOW VOLTAGE / LOW NOISE HEADSTAGE
(Standard SEC SYSTEMS without Adapted Amplifier)
This headstage has an output current range of
15nA into maximum 100 M
, and the noise and bias current
are reduced by a factor of 10 compared to the standard headstage. Therefore, it is recommended for whole
cell patch clamp recordings.
All
current
related signals have to be divided/multiplied by 10.
CURRENT DISPLAY: XXX pA /no decimal point (100 = 100 pA)
HOLDING CURRENT: XXX pA (100 = 100 pA)
Input sensitivity: BNC labeled 1 nA/V has now 0.1 nA/V i.e. 1V = 100 pA
BNC labeled 0.1 nA/V has now 0.01 nA/V i.e. 1V = 10 pA)
Output sensitivity: 1V/nA-100V/nA
BRIDGE BALANCE: (XXX) x10 M
R
EL
mode display: XXXX M
(i.e. 10 is 100 M
)
The
potential
input and output signals are
not
affected.