22
CONTROLS AND
CONNECTORS
FRONT PANEL
ELEMENTS
1 POWER
POWER switch
This switch turns on the power supply. The line
fuse, line voltage selector and power cable
connector are located on the rear panel.
2 VC ERROR
VC ERROR display
This analog display shows the error in the VC
(voltage clamp) mode (command minus recorded
potential). The desired range of operation is
around zero.
3
INTEGRATOR TIME CONSTANT
INTEGRATOR TIME CONSTANT
control and "on/off" switch
·
If the control is turned clockwise the
INTEGRATOR part of the PI (proportional-
integral) control is turned on.
·
The VC controller then has a PI characteristic
(proportional and integral), which reduces the
error considerably (theoretically to zero).
·
The time constant of the integrator is set with
the potentiometer
3.
·
Clockwise: time constant is decreased, effect
of integrator is increased.
·
When using the integrator, step commands
applied to the
VOLTAGE COMMAND
INPUT BNCs (30 / 32)
can cause
overshoots, which can be reduced with the
RISE TIME
control
24
.
4 POTENTIAL FILTER
POTENTIAL FILTER switch
(20Hz – 20 kHz, 16 position rotary switch)
·
2 pole or 4-pole tunable Bessel filter (12 or 24
dB/octave) with 16 corner frequencies for the
POTENTIAL OUTPUT (see also
CURRENT
FILTER 8.
·
The following 16 frequencies can be set:
20, 50, 100, 200, 300, 500, 700, 1k, 1.3k, 2k,
3k, 5k, 8k, 10k, 13k, 20k (Hz)
·
The position of this switch is monitored at
the
FREQUENCY MONITOR
BNC
25
(-8...+7 Volt, 1 Volt / switch position).
Summary of Contents for SEC-05LX
Page 5: ......
Page 6: ......
Page 7: ......
Page 8: ......
Page 9: ......
Page 10: ......
Page 11: ......
Page 12: ......
Page 13: ......
Page 14: ......
Page 15: ......
Page 16: ......
Page 17: ......
Page 18: ......
Page 19: ......
Page 20: ......
Page 21: ......
Page 22: ......
Page 23: ......
Page 24: ......