26
24 RISE TIME
RISE TIME control
Sometimes it is necessary to limit the rise time of
a voltage clamp pulse especially in connection
with PI-controllers to avoid overshooting of the
potential. See also
3
.
25 FREQUENCY MONITOR
FREQUENCY MONITOR BNC connector
The position of the
POTENTIAL FILTER
switch 4
is monitored at the
FREQUENCY
MONITOR
BNC (-8...+7 Volt, 1 Volt / switch
position).
26 MODE SELEC (TTL)
MODE SELECT BNC connector
If the
EXTERNAL CONTROL MODE
. is set
with the
MODE of OPERATION
switch
20
the
mode of operation can be preset by a TTL pulse
applied to the
MODE SELECT BNC connector
26.
A LOW signal presets C mode (switched
current clamp mode), a HIGH signal presets V
mode (switched voltage clamp mode).
See also
MODE OF OPERATION 20
.
27 R
EL
-Mode Switch and
R
EL
mode switch
M
Ω
Ω
LED (light emitting diode)
This switch activates the automatic electrode
resistance measurement mode.
·
In this mode brief current pulses are applied
to the microelectrode and the resistance is
displayed on the upper display in M
W
.
·
The activation of this mode is indicated by
the
M
Ω
Ω
LED
above the switch. The
electrode resistance is measured correctly
regardless of DC offsets.
·
Important
: The displayed resistance may be
dependent on the correct setting of the
capacity compensation (see
28
)
28 CAPACITY COMPENSATION
CAPACITY COMPENSATION
Fine adjustment for the compensation of input
capacitance (ten turn potentiometer, clockwise),
up to 20 pF can be compensated. (Coarse
compensation control on headstage (CC) and
tuning procedure see CAPACITY
COMPENSATION chapter).
Summary of Contents for SEC-05LX
Page 5: ......
Page 6: ......
Page 7: ......
Page 8: ......
Page 9: ......
Page 10: ......
Page 11: ......
Page 12: ......
Page 13: ......
Page 14: ......
Page 15: ......
Page 16: ......
Page 17: ......
Page 18: ......
Page 19: ......
Page 20: ......
Page 21: ......
Page 22: ......
Page 23: ......
Page 24: ......