Section 15
Connecting to other Equipment
70
Model 520A Service Manual
5693-90-01
PRELIMINARY
Output Module in place. The pinouts of the 9-pin analog and 25-pin RS232C
connector are shown below.
Analog Output Setup
15.6
A
CALIBRATE RECORDER
menu within the Model 520A software allows the user
to easily calibrate analog recorders to the voltage levels produced by the Analog
Output Module. To access this feature:
1. With the monitor in normal running mode press and hold the
MENU
key until
SPO2 SETUP OPTIONS
appears on the display. Press the
NEXT
key until
Figure 17.
Analog Output Module
Figure 18.
Analog Output Module Pinouts
1
5
6
9
PR
GNDI
PLETH
SAT
Pin 1
Pulse Rate, 4mV/bpm, 0-1V
Pin 2-3
GNDI (Interface ground)
Pin 4-5
Saturation, 10mV/%, 0-1V
Pin 8-9
Plethysmogram, 0-1V
(Auto Gain Control)
1
13
14
25
RX
TX
CTS
GNDI
VDI
DTR
GNDI
Pin 2
RX
(Receive In)
Pin 3
TX
(Transmit Out)
Pin 6
CTS
(Clear To Send (input))
Pin 7
GNDI (Interface ground)
Pin 20
DTR (Data Transmit Ready
(output))
Pin 21-22
GNDI (Interface ground)
Pin 24-25
VDI
(Int5 volts)
Summary of Contents for OXYPLETH 520A
Page 10: ...Section List of Tables ix Model 520A Service Manual 5693 90 01 PRELIMINARY...
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ......
Page 108: ......
Page 109: ......
Page 110: ......
Page 111: ......
Page 112: ......
Page 113: ......
Page 114: ......
Page 115: ......
Page 116: ......
Page 117: ......
Page 118: ......
Page 119: ......
Page 120: ......
Page 121: ......
Page 122: ......
Page 123: ......
Page 124: ......
Page 125: ......
Page 126: ......
Page 127: ......
Page 128: ......
Page 129: ......
Page 130: ......
Page 131: ......
Page 132: ......
Page 133: ......