Section 15
Connecting to other Equipment
68
Model 520A Service Manual
5693-90-01
PRELIMINARY
2. Connect the printer interface cable, PN:4913-00, to the RS232C connector
1
on the Model 520A. Connect the other end of the cable to the Model 315.
3. Connect the printer’s AC adapter (if desired).
4.
Turn the printer on first
, then turn the Model 520A on.
5. Select
315
from the
Ser.
(Serial) portion of the Model 520A Options Menu.
6. Refer to the Model 315 documentation for further printer details.
Connecting Seiko DPU-411 Thermal Printer
15.4
To connect the SIEKO Model DPU-411 Thermal Printer to the Model 520A:
1. Set the DIP switches located on the bottom of the DPU-411.
First bank of eight switches; 1,2 down (off), 3,4 up (on), 5,6 down (off), 7,8
up (on).
Second bank of six switches; 1-3 up (on), 4-6 down (off).
Figure 14.
Connecting the Model 315 Printer
Figure 15.
Configuration of Seiko Thermal Printer.
MODEL 315 REAR PANEL
POWER SWITCH
AC ADAPTER
4913-00 CABLE
MODEL 520A REAR PANEL
RS232-C INTERFACE
CONNECTOR
DIP SWITCHES
1 2 3 4 5 6 7 8
Seiko DPU-411 bottom view
Summary of Contents for OXYPLETH 520A
Page 10: ...Section List of Tables ix Model 520A Service Manual 5693 90 01 PRELIMINARY...
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ......
Page 108: ......
Page 109: ......
Page 110: ......
Page 111: ......
Page 112: ......
Page 113: ......
Page 114: ......
Page 115: ......
Page 116: ......
Page 117: ......
Page 118: ......
Page 119: ......
Page 120: ......
Page 121: ......
Page 122: ......
Page 123: ......
Page 124: ......
Page 125: ......
Page 126: ......
Page 127: ......
Page 128: ......
Page 129: ......
Page 130: ......
Page 131: ......
Page 132: ......
Page 133: ......