Model 520A Service Manual
49
PRELIMINARY
PROBE FAULTY RD
Sensor faulty. Remove sensor from use and contact qualified
service personnel.
☞
Sensor Red LED has failed or the cable connections are
open or shorted.
PROBE FAULTY IR
Sensor faulty. Remove sensor from use and contact qualified
service personnel.
☞
Sensor Red LED has failed or the cable connections are
open or shorted.
PROBE OFF PATIENT
Sensor disconnected from patient, improperly applied, or
placed on an area too translucent for proper sensor operation.
Reposition sensor.
PULSE-HIGH
Selected pulse rate high alert limit has been violated.
PULSE-LOW
Selected pulse rate low alert limit has been violated.
PULSE OUT OF RANGE
Pulse rate is less than 30 bpm or is greater than 250 bpm.
Revision 2.0
APR/28/92
Monitor software revision level. Displayed when monitor is
turned on while pressing
AUDIO
key.
SpO2-HIGH
Selected saturation high alert limit has been violated.
SpO2-LOW
Selected saturation low alert limit has been violated.
Message Center Display
Possible Explanation
Summary of Contents for OXYPLETH 520A
Page 10: ...Section List of Tables ix Model 520A Service Manual 5693 90 01 PRELIMINARY...
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ......
Page 108: ......
Page 109: ......
Page 110: ......
Page 111: ......
Page 112: ......
Page 113: ......
Page 114: ......
Page 115: ......
Page 116: ......
Page 117: ......
Page 118: ......
Page 119: ......
Page 120: ......
Page 121: ......
Page 122: ......
Page 123: ......
Page 124: ......
Page 125: ......
Page 126: ......
Page 127: ......
Page 128: ......
Page 129: ......
Page 130: ......
Page 131: ......
Page 132: ......
Page 133: ......