Assembly Exchanges
5-Jun-01
Model 520A Service Manual
41
PRELIMINARY
3. Using needle-nosed pliers, pull the Voltage Selector Drum from the PEM.
Note the orientation of the drum; the proper voltage should face out.
4. Set the Voltage Selector Drum so that the printed side of the correct voltage
faces you. The voltage selections are pictured below.
5. Snap the Voltage Selector Drum back into the PEM. Be sure that the Voltage
Setting Indicator shows the proper voltage. Close the Fuse Access Door.
Assembly Exchanges
10.6
Disassembly should be performed by qualified personnel. Follow proper grounding
procedures to avoid damage to internal components from static discharge.
Figure 6.
Power Entry Module Voltage Selector Drum Removal
Figure 7.
Power Entry Module Voltage Selection Adjustment
115V
ac
O
Voltage Selector Drum
115V
ac
O
OR
115V
ac
230V
ac
Summary of Contents for OXYPLETH 520A
Page 10: ...Section List of Tables ix Model 520A Service Manual 5693 90 01 PRELIMINARY...
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ......
Page 108: ......
Page 109: ......
Page 110: ......
Page 111: ......
Page 112: ......
Page 113: ......
Page 114: ......
Page 115: ......
Page 116: ......
Page 117: ......
Page 118: ......
Page 119: ......
Page 120: ......
Page 121: ......
Page 122: ......
Page 123: ......
Page 124: ......
Page 125: ......
Page 126: ......
Page 127: ......
Page 128: ......
Page 129: ......
Page 130: ......
Page 131: ......
Page 132: ......
Page 133: ......