Norsonic Nor150
Instruction Manual
•
Repeat
, which causes the instrument to store
the acquired data and then restart immediately
and make another measurement using the same
measurement setup and duration. Repeat applies
to measurements terminated by themselves only.
If you terminate a measurement by pressing stop,
the instrument will not restart. There is no time gap
between the measurements.
Synchronized – an example
Assume that you set up the instrument to measure in
periods of one hour and that you start the measurement
at 08:52:40. The first period will last 7 minutes and 20
seconds ending at 09:00:00. The next measurement
will be started at 09:00:00, followed by a new
measurement each full hour.
We recommend limiting the use of the synchronized
feature for measurement periods which are either a
multiple of one hour or one hour divided by a whole
number as the feature has been designed with this
limitation in mind.
If you want to apply short measurement periods below
a few minutes, we generally recommend using the level
versus time feature instead (time profile), since this gives
you a much more flexible way of displaying and work
with the stored data.
Figure 9.3
Weather Station
Turns the weater station logging on/off.
A submenu (figure 9.3) is avalable where you can select
which parameter to log. Supported weatherstation is
Thies Clima US model 4.9200.00.000.
The data logged from the weather station is not visible
on the Nor145 display but is stored along with the
measurement data for later analysis in NorReview or
NorCloud.