179
Norsonic Nor150
Instruction Manual
Measurement range for C-weighted peak
levels
Frequency
31.5 Hz 1 kHz 4 kHz 8 kHz 12.5 kHz
Upper level
137
140
139
137
134
Lower level
45
45
45
45
45
Reference level
114
114
114
114
114
Levels above the “Upper level” will normally indicate
Overload.
The Nor150 used for electrical
measurements
By replacing the microphone with an input adapter
(Nor1447) the instrument is well suited for electrical
measurements. Alternatively the complete transducer
can be removed and replaced with the BNC to Lemo
cable type Nor1438.
Please note that when the Nor1438 cable is used it is
needed to specify a new transducer with a new name
in the sensor set up menu system. Then the instrument
can also be configured to feed an IEPE device.
Electrical verification
measurements
Electrical signals resembling the microphone signal
is used to assess conformance to the specifications
and the sound level meter standard. The input adapter
Nor1447 with an internal capacity of 18 pF (± 20%) is
then used for feeding the signal into the instrument.
The reference level is 114 dB SPL (10 Pa) and the nominal
microphone sensitivity for Nor1225 is 50 mV/ Pa, so the
reference level shall be indicated with an input voltage
level of 500 mV ± 3 dB.
Maximum allowed input signal for this purpose is 10 V
rms that is equivalent to 140 dB SPL.
Power supply
After the instrument is turned on, it takes some time to
boot up correctly. In addition to this all electronic parts
(especially the 200 V polarization voltage) need some
time to stabilize properly. It takes two minutes from the
power is turned on until the instrument is ready for use.
Internal battery
Battery type:
Li-Ion package (simply called Nor150/
Battery) with power capacity control circuit. Each
battery pack has a “fuel gauge” and its own serial
number. The battery and the instrument are designed
so the full battery capacity will last for a normal 8 hour
work day.
Voltage:
7.4 V
Capacity:
3760 mAh
Charging time using Nor345A mains adapter:
4 hours (80% in 2 hours)
No other battery type than Nor150/Battery from
Norsonic can be used in the instrument. Replacement
or extra batteries are available from all Norsonic,
representatives.
The battery must be treated with care. Never store in a
hot place or in direct sunlight. Keep free from moisture
and dust.
Recycling:
These types of batteries are poisonous.
Never waste the battery without knowing that it will be
handled as required. If you are in doubt, return it to the
Norsonic representative.
Summary of Contents for NOR150
Page 4: ......
Page 14: ...x Contents ...
Page 15: ...xi Norsonic Nor150 Instruction Manual nor150 SOUND VIBRATION ANALYSER ...
Page 16: ......
Page 204: ......
Page 205: ......
Page 206: ......