3
Norsonic Nor150
Instruction Manual
Taking a closer look at
the instrument
Be sure to take utmost care when mounting a
microphone cartridge onto a preamplifier. To avoid
electrical shock from the 200 V polarization voltage
always keep the preamplifier disconnected whenever
you are mounting a cartridge onto a preamplifier and
screw finger tight only!
The picture shows the Nor150 fitted with the standard
preamplifier Nor1209 and the microphone Nor1225
mounted in sound channel one.
The instrument is powered from an internal rechargeable
Li-ion battery pack. The battery is fully charged when
leaving the factory, but due to self-discharge, you may
experience the battery gauge shows a lower value than
full. The batteries will be charged once the instrument
is connected to mains via the supplied mains adapter
Nor345A/Nor345B.
Always keep the microphone cartridge mounted on
the preamplifier. This is the safest way to avoid
damage and access of dirt on the contact pin
between the preamplifier and microphone cartridge.
The preamplifier input has extremely high input
impedance (10 Giga ohms) in order not to load the
microphone cartridge. Hence, dust, finger prints or
other types of contamination may change the
sensitivity of the microphone, especially at high
humidity.