Chapter 1
Important information
Caution!
Never expose the batteries to temperatures
outside maximum operation temperature range;
-10 °C / 14 °F to +50 °C / 122 °F.
Maximum storage temperature range is -30 °C /
-22 °F to +50 °C / 122 °F. Room temperature is
recommended for long time storage. Always fully
charge the batteries before storage. Recharge
batteries fully every third month for long time
storage.
Never use mechanical damaged batteries.
Shipment of batteries via air or brought as hand
luggage / luggage are governed by IATA /
UN3481, PI 967 section II when shipped as a
part of the instrument and PI 966 part II if shipped
separate. See IATA standard for labelling the
parcel and to acquire latest information about
shipment instructions.
NOTE Damaged batteries are strictly prohibited to
be shipped as airfreight or brought onto a plain as
hand luggage or checked in luggage.
Precautions
You probably already know it but…
…. always turn on the password protection if you
connect the instrument to a public or unsecured
network. If not, the instrument is an easy victim for
cyber-attack. It is a good practice to always use a FTP
password regardless of the connection environment.
Also, enable a password protection for the NorRemote
program.
“Security” on page 146
… the microphone is an especially fragile device. It is
easily broken, so take care.
…. always keep the microphone cartridge mounted on
the preamplifier. This is the safest way to avoid damage
and access of dirt on the contact pin between the
preamplifier and microphone cartridge.
… keep the instrument in its carrying case, don’t leave
it everywhere.
… even the instrument is a field instrument, prevent it
from direct contact with dust and moisture.
… Nor150 is a measurement instrument; protect it from
impacts and strong vibration.
… never store the instrument with empty batteries. This
may permanent damage the batteries
… fully recharge batteries every month if the instrument
is not in use.
… always calibrate the microphone before and after a
measurement.
… send the instrument for verification at an accredited
laboratory minimum every 24 months.