CONTROL settings
Host IP address
To help prevent unauthorized access, the Koheras ACOUSTIK can be configured
to only accept packets from a single IP address assigned to the management PC.
Configure the
Host IP address
to the IP address of your management PC. When
set, the ACOUSTIK only accepts packets with a source address that matches the
Host IP address
set and ignores all others. When set to 000.000.000.000
(default setting), the ACOUSTIK accepts packets with any source IP address.
Host port
The
Host port
sets the port address the ACOUSTIK uses for transmission of TCP
or UDP packets. The
Host Port
address set in the network connection of
CONTROL must match this address – default: 0.
Note:
If the
Host port
is set to 0, the ACOUSTIK will use the same port address
for transmission as for reception i.e. the
System port
setting. The ports addresses
set in ACOUSTIK and CONTROL must match.
MAC Address
A unique
MAC address
of the ACOUSTIK is displayed only and cannot be set.
Watchdog
The Koheras ACOUSTIK features a watchdog for monitoring of USB or Ethernet
communication. If CONTROL is disconnected from the system, the watchdog
disables emission on each module after a set timeout expires.
Enable Watchdog
To turn the watchdog on, select ON from the drop-down menu next to Enable
watchdog. To turn the watchdog off, select OFF.
Watchdog timeout
When CONTROL communications is disconnected, the Watchdog timer starts.
When it reaches the Watchdog timeout setting, the emission from each
ACOUSTIK module is disabled. The Watchdog feature must be set to ON, then
set the timeout value from 1 to 255 seconds using the text input box.
Figure 35
Watchdog setting
Clock
Within the Clock setting menu the system date and time can be read or set
automatically by updating the settings from a connected PC.
When ON, laser emission
is disabled after timeout expiry
Summary of Contents for Koheras ACOUSTIK
Page 1: ...Item 800 633 01 Koheras ACOUSTIK PRODUCT GUIDE Multi Channel Modular Laser System...
Page 10: ...10...
Page 14: ...14...
Page 16: ...16...
Page 26: ...System labels 26...
Page 28: ...28...
Page 34: ...Operating and storage environment 34...
Page 42: ...42...
Page 78: ...Module overview 78...
Page 80: ...80...
Page 82: ...82 Figure 44 Mechanical dimensions 483 00 45 00 302 50 4 50 125 00 2...
Page 92: ...Installing CONTROL 92...
Page 95: ......
Page 96: ...1 Koheras ACOUSTIK Product Guide Revision 1 0 09 2021 W 10456...