Function (Fct.)
Input Range
Description
2.1.6.1
ADDRESS
Enter value between 0 to 15
Sets the address of the device when this latter is connected to HART
Multidrop networks.
The current output drifts to 4 mA.
0 = 4 – 20 mA output current active
1 – 15 = in multidrop mode
0
2.1.7.0
KONVERZIÓS
TÁBLÁZAT
2.1.7.1 VOLUME
UNIT
Select:
m
3
, l, US Gal, Ft
3
, bbl, m
3
/h,
Ft
3
/h, kg, Metric Tons, US Tons
Unit for conversion values (“volume table”).
The selected unit is used to define the strapping table values.
Litre [l]
2.1.7.2
INPUT TABLE
Between 0 to 20 points
This function defines the strapping table.
The maximum number of points is 20.
Each subsequent value must be greater than the previous one.
The length and volume units can be changed later without affecting the
settings in the table. The units of length and volume used in Fct.s 1.2.4 and
1.2.5 shall also apply here.
0
(i.e. no volume table)
2.1.7.3
DELETE TABLE
This function deletes the strapping table.
2.2.0 TESTS
2.2.1 TEST
OUTPUT
Select:
4 mA, 8 mA, 12 mA, 20 mA
This function allows the current output to be tested. The output can be set
to one of the listed values. Using a reference ammeter, the calibration of
the current output can be verified.
2.3.0 SERVICE
Restricted access to the factory configuration menu. These parameters
may be accessed via Fct. 3.2 “Specialist PSW (password).
3.0
ACCESS RIGHTS
3.1
MAINTENANCE
PSW
Yes or No.
Enter 9-character code if “YES”.
Disables the access lock on the configuration menu.
The password must contain exactly 9 characters.
E, R or U are used only.
The password is displayed in a scrambled format. It allows NIVELCO to
decode the password in case it was forgotten.
No
3.2
SPECIALIST PSW
Contact a specialist for Service Parameters.