4.
MECHANICAL INSTALLATION
4.1.
H
ANDLING AND STORAGE
When manually carrying, hold it with both hands, or when powerlifted, pay attention to
handle the device with care, lifting it by the converter housing. If necessary, use lifting
gear. No attempt should be made to lift the instrument by its probe.
Caution: The probe is a critical gauge component. Do not damage – Handle with care!
Avoiding blows – avoid hard blows, jolts, impacts, etc.
Caution: fragile electronics
Avoid bending (single rod and coaxial probes) – Support the probe to avoid bending.
Avoid cable kinks and fraying
Do not coil the cable with less than 400 mm (16 inch) in diameter. Cable kinks
or fraying will cause measurement errors.
Storage temperature