5.2.2.
MicroTREK’s behaviour in manual programming mode
After power-up MicroTREK shows the measured value on the SAP-300 display (referred as “display”).
REMOTE MODE:
If the instrument senses external HART
communication it changes its display mode and shows a "REMOTE MODE" message in the bottom of the display.
In this mode the measured values are refreshed according to the queries of the external HART
master.
If the HART
master does not refresh the display, the display will show the last measured values.
In the absence of SAP-300, the COM LED indicates the HART
communication.
If HART
communication stops, the COM LED is turned off after 120 sec.
Press
E
button to enter the menu. Use and buttons to navigate in the menu.
Enter the selected menu point with
E
button. Exit to the previous menu with button.
Buttons are working only in the presence of the SAP-300 module!
When leaving the MicroTREK in (programming) menu after 30 minutes the instrument automatically returns to measuring mode.
If SAP-300 is removed the instrument instantly returns to measuring mode.
Because of manual programming (with SAP-300) and remote programming (with external HART
master, MultiCONT or EView2 software) cannot be performed at the
same time (as both of them act like a HART
master) only one programming mode has priority and this is manual programming.
During manual programming the instrument sends the "BUSY" response to the external HART
master device. (HART
Response code: 32 – Device is busy)