Code setting
address
Item
Code
Description of code
C36
Last paratmeter
displayed in circle
display
00~05
Same as above
C40
The date sequence of
IEEE formate float of
Modbus
01
Lower 16bits address in front, the lower 8bits in the 16bits
address in front
02
Higher 16bits address in front, the Higher 8bits in the
16bits address in front
03
Lower 16bits address in front, the higher 8bits in the 16bits
address in front
04
Higher 16bits address in front, the lower 8bits in the 16bits
address in front
C41
Shape of pipeline
00
Round pipeline(D:D010)
01
Rectangel pipeline(L:D018 W:D019)
C47
Set password
00
Keep the password
01
Change the pass word
C49
Status
00
Working status
01
4~20mA output calibration statues
02
Flow rate calibration statues
03
Temperature calibration statues
04
Pressure calibration statues
C50
Total flow reset
00
Reset total flow to 0
01
Default
C60
Restore to backup
date
06
Restore to backup date
C61
Save setting backup
16
Save current setting for backup
C80
Version No. of
hardware
Read Only
C81
Version No. of
software
Read Only
C82
Calibration date
Read Only
C83
Number of errors
Read Only
C84
Version No. of
HART
Read Only
C86
Communication
interface
00
Modbus Read Only
01
Hart Read Only
C87
Pressure sensor type
00
No pressure sensor Read Only
01
Piezo Read Only
02
Piezo Read Only
03
4~20mA Read Only
04
0~10mA Read Only
C88
Product ID No.
Read Only
Note:
1) If the unit of flow rate is changed or measurement changed from flow rate to mass flow, users can
reset the total flow to 0 or record the current total flow, as the number of total flow will not change
according to unit.
2) Total flow can only be output by pulse, temperature and pressure can only be output by frequency