100
LAN COMMUNICATION
:
Data Format
MLM26*RTZ002F
●
LM data
Tag name
Description
Data type
Necessity
Example
<Measure Type="LM">
Measurement type tag
-
○
<MeasureMode>
Measurement mode
String (64 characters)
(AutoNormal, Normal,
AutoProgressive,
Progressive, Contact,
Layout, Convex,
Index, and Other only)
△
AutoNormal
<DiopterStep unit="D">
Diopter unit increment
Double
△
0.25
<AxisStep unit="deg">
Cylinder axis unit increment
int
△
1
<CylinderMode>
Cylinder mode
String (+, -, and
+/-
[MIX]
only)
△
-
<PrismDiopterStep unit="pri">
Prism power unit increment
Double
△
0.25
<PrismBaseStep unit="deg">
Prism base direction unit increment
int
△
1
<PrismMode>
Prism mode
String (pb and xy only)
△
xy
<AddMode>
Add power moder
String (add and nsph
only)
△
add
<LM>
LM measurement data
-
( )
<S>
Single state measurement value
-
( )
<Sphere unit="D">
Spherical power (D)
Double
○
-3.00
<Cylinder unit="D">
Cylindrical power (D)
Double
○
-1.00
<Axis unit="deg">
Cylinder axis (°)
int
○
180
<SE unit="D">
Spherical equivalent value (D)
Double
△
-3.50
<ADD unit="D">
1st add power (D)
Double
△
1.00
<ADD2 unit="D">
2nd add power (D)
Double
△
1.50
<NearSphere unit="D">
1st near power (D)
Double
△
-2.00
<NearSphere2 unit="D">
2nd near power (D)
Double
△
-1.50
<Prism unit="pri">
Prism power (
△
)
Double
△
0.50
<PrismBase unit="deg">
Prism base direction (°)
int
△
180
<PrismX unit="pri" base="out">
Horizontal prism power
(
△
)
(base in and base out only)
Double
△
0.50
<PrismY unit="pri" base="up">
Vertical prism power (
△
)
(base up and base down only)
Double
△
0.00
<UVTransmittance unit="%">
UV transmittance (%)
int
△
5
<R>
Right-eye lens measurement value
-
( )
<Sphere unit="D">
Spherical power (D)
Double
○
-3.00
<Cylinder unit="D">
Cylindrical power (D)
Double
○
-1.00
<Axis unit="deg">
Cylinder axis (°)
int
○
180
<SE unit="D">
Spherical equivalent value (D)
Double
△
-3.50
<ADD unit="D">
1st add power (D)
Double
△
1.00
<ADD2 unit="D">
2nd add power (D)
Double
△
1.50
<NearSphere unit="D">
1st near power (D)
Double
△
-2.00
<NearSphere2 unit="D">
2nd near power (D)
Double
△
-1.50
<Prism unit="pri">
Prism power (
△
)
Double
△
0.50
<PrismBase unit="deg">
Prism base direction (°)
int
△
180
Summary of Contents for LM-1800P
Page 1: ...INTERFACE MANUAL MLM26 RTZ002F LM 1800P LM 1800PD AUTO LENSMETER...
Page 6: ...2 OUTLINE OF COMMUNICATION Outline of Communication Function MLM26 RTZ002F...
Page 54: ...50 RS 232C Communication and USB Communication Example of Data Output MLM26 RTZ002F...
Page 70: ...66 COMMUNICATION PROCEDURE other than NCP20 Flowchart MLM26 RTZ002F...