51
MLM26*RTZ002F
3
3.
COMMUNICATION PROCEDURE
(other than NCP20)
3.1
Outline
The procedure for transmitting data measured by the LM to the computer is explained.
1
Save the measurement data on the LM.
Press the read key of the LM or send “RD command” from the computer.
2
Transmit the saved data to the computer.
Press the print button
(or export button
) or send “SD command” from the computer.
3
As the computer receives the measurement data, it displays the data on the
screen or saves it to a floppy disk.
Therefore, the computer needs a program that sends or receives these “commands”
or “data”, and analyzes, displays or saves the received “data”.
In this manner, the operator can choose the following options for data transmission and reception: To
perform automatic data transmission with the operation buttons on the LM, to use the computer to give
all the commands to the LM from the data storage to data transmission, and to use the operation but-
tons on the LM and the computer in combination.
• The icon indication differs depending on the parameter settings. For details, refer to each
operator’s manual.