40
RS-232C Communication and USB Communication
:
Example of Data Output
MLM26*RTZ002F
(11)In the case of the maximum data size in normal measurement mode (LM-1800PD)
Measurement conditions:
Trifocal lens measurement with the left- and right- eye lenses designated
Right: SPH = -1.25 D, CYL = -0.75 D, AXIS = 120
º
ADD = +2.00 D, ADD2 = +3.00 D
Representation form of near portion power = NEAR SPH NSPH = +0.75
D, NSPH2 = +1.75 D
PRISM = IN 2.25
Δ
, DOWN 2.00
Δ
(BU/D BI/O setting)
Transmittance measurement Green = 80%, UV = 20%
RPD = 31.5 mm
Left: SPH = -2.00 D, CYL = -0.50 D, AXIS = 180
º
ADD = +2.25 D, ADD2 = +3.50 D
Representation form of near portion power = NEAR SPH NSPH = +0.25
D, NSPH2 = +1.50 D
PRISM = OUT 1.25
Δ
, UP 2.00
Δ
(BU/D BI/O setting)
Transmittance measurement Green = 80%, UV = 20%
LPD = 32.5 mm
Total PD = 64.0 mm
Net prism measurement Net H. Pri = IN 1.25
Δ
, Net V. Pri = UP1.00
Δ
CR code = Added
• This is an example of LM-1800PD of the maximum data size (300 characters) in normal
lens measurement mode.
• When the Near parameter is set to “Near Sph”, both the addition powers (ADD) and near
spherical data (NSPH) are output.
• When the Net prism parameter is set to “On”, the difference between the right-eye and left-
eye lenses are output after measuring net prism as a net prism value.
• When reading two kinds of barcode data: Operator/Patient ID.
1
5 (Characters)
SOH
D
L
M
STX
→
To be continued
01H 44H 4CH 4DH 02H
10
15
20
I
D
N
I
D
E
K
/
L
M
-
1
8
0
0
P
d
ETB CR
→
To be
continued
49H 44H 4EH 49H 44H 45H 4BH 2FH 4CH 4DH 2DH 31H 38H 30H 30H 50H 64H 17H 0DH
Company name/ Model name (LM-1800PD)
25
30
35
40
I
P
1
2
3
4
5
6
7
8
9
0
1
2
3
4
5
6 ETB CR
→
To be
continued
49H 50H 31H 32H 33H 34H 35H 36H 37H 38H 39H 30H 31H 32H 33H 34H 35H 36H 17H 0DH
Barcode data 1 (Up to 16 characters)
Summary of Contents for LM-1800P
Page 1: ...INTERFACE MANUAL MLM26 RTZ002F LM 1800P LM 1800PD AUTO LENSMETER...
Page 6: ...2 OUTLINE OF COMMUNICATION Outline of Communication Function MLM26 RTZ002F...
Page 54: ...50 RS 232C Communication and USB Communication Example of Data Output MLM26 RTZ002F...
Page 70: ...66 COMMUNICATION PROCEDURE other than NCP20 Flowchart MLM26 RTZ002F...