26
RS-232C Communication and USB Communication
:
Example of Data Output
MLM26*RTZ002F
(3) In the case of single data in contact lens measurement mode:
Measurement conditions:
Contact lens measurement with the single state designated
SPH = +2.00 D, CYL = +0.50 D, AXIS = 60°, SE = +2.25 D
PRISM = 1.25
Δ
, BASE = 70° (P-B setting)
No transmittance measurement performed
CR code = Added
Operator/Patient ID = None (No barcode reading performed)
• In the contact lens measurement mode, the SE value is added to the end of SPH, CYL and
AXIS data.
• The prism representation mode is different from case (2).
• The “CR code” is added to the end of “ETB” and “EOT”.
SOH
D
L
M
STX
→
To be continued
01H 44H 4CH 4DH 02H
I
D
N
I
D
E
K
/
L
M
-
1
8
0
0
P
d
ETB CR
→
To be
continued
49H 44H 4EH 49H 44H 45H 4BH 2FH 4CH 4DH 2DH 31H 38H 30H 30H 50H 64H 17H 0DH
Company name/ Model name (LM-1800PD)
D
A
2
0
1
0
.
1
1
.
2
3
.
1
5
:
3
0
ETB CR
→
To be
continued
44H 41H 32H 30H 31H 30H 2EH 31H 31H 2EH 32H 33H 2EH 31H 35H 3AH 33H 30H 17H 0DH
Year
Month
Day
Hour
Minute
+
0
2
.
0
0
+
0
0
.
5
0
0
6
0
ETB CR
→
To be
continued
20H 20H 2BH 30H 32H 2EH 30H 30H 2BH 30H 30H 2EH 35H 30H 30H 36H 30H 17H 0DH
Single SPH
Single CYL
Single AXIS
S
+
0
2
.
2
5
ETB CR
P
0
1
.
2
5
ETB CR
→
To be
continued
53H 20H 2BH 30H 32H 2EH 32H 35H 17H 0DH 50H 20H 30H 31H 2EH 32H 35H 17H 0DH
Single SE
Single PRISM
B
0
7
0
ETB CR EOT CR
42H 20H 30H 37H 30H 17H 0DH 04H 0DH
Single BASE
Summary of Contents for LM-1800P
Page 1: ...INTERFACE MANUAL MLM26 RTZ002F LM 1800P LM 1800PD AUTO LENSMETER...
Page 6: ...2 OUTLINE OF COMMUNICATION Outline of Communication Function MLM26 RTZ002F...
Page 54: ...50 RS 232C Communication and USB Communication Example of Data Output MLM26 RTZ002F...
Page 70: ...66 COMMUNICATION PROCEDURE other than NCP20 Flowchart MLM26 RTZ002F...