![Nidek Medical LE-700 Installation Manual Download Page 66](http://html1.mh-extra.com/html/nidek-medical/le-700/le-700_installation-manual_1688019066.webp)
60
CHECK AND ADJUSTMENT PROCEDURES: Polish Size Check and Adjustment
9
Specify the polishing conditions.
1)
Set each item as follows:
•
For flat edge polishing
“Material”
→
CR39
“Type”
→
Single
“Frame”
→
Flat
“Soft”
→
None
“Polish”
→
Exec
“Layout”
→
Passive
“FC”
→
None
•
For bevel polishing
“Material”
→
CR39
“Type”
→
Single
“Frame”
→
Metal
“Bevel position”
→
Auto
“Soft”
→
None
“Polish”
→
Exec
“Layout”
→
Passive
“FC”
→
None
2)
Press .
The layout screen appears.
10
Continue the processing method on the layout screen using the same procedures
explained by the following steps starting from size setting.
to
11
Measure the outer diameter of the processed lens to confirm the value is within the
following tolerance.
“3.11.1 Polish size check” (page 56)
12
until the outer diameter is within the tolerance of
φ
45.0±0.1 mm.