51
CHECK AND ADJUSTMENT PROCEDURES: Polish Axis Check and Adjustment
3.10 Polish Axis Check and Adjustment
Check the edge surface of a plastic lens flat edge pol-
ished using the Square 45 internal data. If the right or
left of the lens edge remains unpolished, change the
Polish axis parameter.
3.10.1 Polish axis check
1
Enter service mode from the LE-700 home screen.
“3.2.1 Entering service mode” (page 20)
2
Press the “Maintenance” tab.
The maintenance screen appears.
3
Press Square 45
.
The processing condition screen appears.
4
Specify the flat edge polishing conditions.
1)
Set each item as follows:
“Material”
→
CR39
“Type”
→
Single
“Frame”
→
Flat
“Soft”
→
None
“Polish”
→
Exec
“Layout”
→
Passive
“FC”
→
None
2)
Press .
The layout screen appears.
Lens front surface
When the unpolished portion is on
the left side of the lens edge.