58
CHECK AND ADJUSTMENT PROCEDURES: Polish Size Check and Adjustment
3.11.2 Polish size adjustment
1
Enter service mode from the LE-700 home screen.
“3.2.1 Entering service mode” (page 20)
2
Press the “Grinding” tab.
3
Press Tab 5.
Scroll up the scroll bar.
4
Press the numeric fields of “CR39, Bevel”
and “CR39, Flat” next to “Polish size”.
The numeric keypad appears.
5
Enter an adjustment value with the numeric
keypad and change the parameter value.
Change the parameter by the difference between
45.00 mm and the measured outer diameter.
Ex.) When the measured outer diameter is 45.15
mm, decrease the parameter value by 0.15.