CHECK AND ADJUSTMENT PROCEDURES: Service Mode
3.2.2
Loading internal shape data
1
Enter service mode from the LE-700 home screen.
“3.2.1 Entering service mode” (page 20)
2
Press the “Maintenance” tab.
The maintenance screen appears.
3
Load the internal shape data.
1)
Press the desired internal shape button.
Circle 45: Press
.
Square 45: Press
.
The processing condition screen appears.
2)
Select the desired processing condition and
press .
3)
The layout screen appears.
(Example shown to the right: Circle 45 shape
data)
CAUTION
• Rectangle 50 x 25, Process test, and Multirole shape data are not used during installation. Refer to the LE-
700 Service Manual for shape data.