4.2 Sweep
WF1945B
4-33
• Internal trigger source
:
The internal trigger signal is produced at the rate indicated below. Even is set to
less than 100 ms, the trigger is applied only at 100 ms intervals.
INT
100.0ms
F-SWP:
TYPE
SOURCE
MODE
4
• START-STATE and STOP-STATE :
Set output to start and stop values, respectively.
Both START-STATE and STOP-STATE oscillation can be
conducted together with a gated sweep. Since sweep
synchronization output becomes the respective start and stop
states, full-scale adjustments of the recorder and the status of
external equipment can be checked.
"
4.2 Sweep (
■
Sweep value and Z-MARKER/SYNC/
X- DRIVE outputs)", cf.
•
Sweep type set to duty :
Output waveform fixed to squarewave (variable duty) and waveform
(Function) cannot be selected.
During sweep, multiple pulses may be produced within 1 period output
as indicated in the figure.
•
If the frequency is low and the sweep rest period short, in some cases, oscillation may not stop
immediately at the end of sweep.
•
When an amplitude sweep is set with 0 as the start or stop values and the sweep function is set with
LOG
/ LOG
/ LOG
, "SETTING CONFLICT 001" is displayed continuously during a
short trigger cycle and the menu may not be displayed. In such cases, switch the trigger source, or
stop input to the SWEEP PAUSE IN connector.
Trigger rate
Summary of Contents for WF1945B
Page 2: ......
Page 3: ...DA00012505 002 WF1945B MULTIFUNCTION SYNTHESIZER Instruction Manual...
Page 4: ......
Page 8: ......
Page 20: ......
Page 30: ......
Page 62: ......
Page 100: ...4 2 Sweep 4 38 WF1945B Sweep value and Z MARKER SYNC X DRIVE outputs is approx 100 to 125 s...
Page 101: ...4 2 Sweep WF1945B 4 39...
Page 128: ......
Page 158: ......
Page 182: ......
Page 184: ......
Page 186: ......
Page 187: ......