5.1 Convenient Settings
5-4
WF1945B
Squarewave pulse width setting (
ENTRY
→
WIDTH
)
Operation to set the squarewave pulse width is described. The setting changes the width of the pulse
(portion indicated by arrows
).
Select squarewave (
variable duty).
Operation:
①
Press the
ENTRY
key, then the
WIDTH
key to produce the following display.
W 0.000500000 s
A0.1000Vp-p O+0.0000V /OPEN
②
Set the width with the keypad or
MODIFY
dial.
Other:
• The table indicates the effects on other parameters when the frequency, period, width or duty is
changed.
Change
↓
Frequency
(FREQ)
Period
(PERIOD)
Pulse width
(WIDTH)
Duty
(DUTY)
Frequency
(FREQ)
Changed
Changed
Unchanged
Period
(PERIOD)
Changed Unchanged
Changed
Pulse width
(WIDTH)
Unchanged Unchanged
Changed
Duty
(DUTY)
Unchanged Unchanged
Changed
• Due to the oscillation period and duty relationship, the pulse can be lost when the actual pulse width is
less than 25 ns. An error message is displayed at this type of setting.
Also, jitter increases when the pulse width is less than 100 ns. A warning message is displayed at
this type of setting.
Summary of Contents for WF1945B
Page 2: ......
Page 3: ...DA00012505 002 WF1945B MULTIFUNCTION SYNTHESIZER Instruction Manual...
Page 4: ......
Page 8: ......
Page 20: ......
Page 30: ......
Page 62: ......
Page 100: ...4 2 Sweep 4 38 WF1945B Sweep value and Z MARKER SYNC X DRIVE outputs is approx 100 to 125 s...
Page 101: ...4 2 Sweep WF1945B 4 39...
Page 128: ......
Page 158: ......
Page 182: ......
Page 184: ......
Page 186: ......
Page 187: ......