4.1 Burst oscillation
WF1945B
4-11
(3) Set STOP-LEVEL
①
Use the
and
keys to produce the following display (STOP-LEVEL flashes).
OFF
GATE:
TYPE
SOURCE
STOP-LEVEL
②
Press the
ENTER
key.
Turn the
MODIFY
dial to produce the following display (ON flashes).
ON 0.00 %
GATE:
TYPE
SOURCE
STOP-LEVEL
③
Press the
key, then set the stop level with the keypad or
MODIFY
dial.
In this example, the setting is 0 %.
The stop level is the percentage with respect to the maximum positive (+100 %) and negative (-100 %)
amplitude.
④
After setting, press the
EXIT
key to release the setting mode.
The above selects sets burst oscillation (type: gate). Oscillation occurs when a high level signal is
applied to the TRIG/SWEEP IN connector. If the connection is open, oscillation continues due to
internal pullup.
Other settings:
•
At the above settings, when the waveform is squarewave, a three squarewave is obtained such as
shown in the following figure.
• Internal gate source
: A 50 % duty gate signal is generated at the following period for
oscillation start/stop.
INT
H-ON
1.000ms
GATE:
TYPE
SOURCE
STOP-LEVEL
(When stop level is off, the oscillation stops at
either high or low level.)
Summary of Contents for WF1945B
Page 2: ......
Page 3: ...DA00012505 002 WF1945B MULTIFUNCTION SYNTHESIZER Instruction Manual...
Page 4: ......
Page 8: ......
Page 20: ......
Page 30: ......
Page 62: ......
Page 100: ...4 2 Sweep 4 38 WF1945B Sweep value and Z MARKER SYNC X DRIVE outputs is approx 100 to 125 s...
Page 101: ...4 2 Sweep WF1945B 4 39...
Page 128: ......
Page 158: ......
Page 182: ......
Page 184: ......
Page 186: ......
Page 187: ......